Several additional parameters are used to calibrate the Z-axis in closed loop operation:
Z Sensor Sens:The sensitivity, in nm/V, of the optical Z-axis sensor when Z Sensor Gain Factor = 1.
Z Sensor Gain Factor:During the engage, the Z piezo is ramped to find the Z piezo range. The sensor gain is adjusted to set a Z sensor range that is:
(Z sensor range factor)*(Z piezo range)=gain of the Z sensor
Make this larger than 1 to ensure that the dynamic range is covered. Note that the resolution of the Z sensor ADC is decreased by this amount.
Z CL Range Factor:The range of the Z sensor that is used in Z closed loop mode (ramping). Making this larger increases the chance that part of the ramp may reach the end of the scanner range.